How to use a seawater refractometer
D-D’s official ATC seawater refractometer calibration and readings, Milwaukee’s MA887 digital steps, and Brightwell’s 35 ppt calibration standard.
A seawater refractometer measures salinity (and often specific gravity) from a few drops of water. This guide follows D-D’s official ATC seawater refractometer instructions, Milwaukee’s official MA887 instruction manual, and Brightwell Aquatics’ official calibration-standard directions. It is about how to calibrate and take a reading, not about which salt mix to buy — see Which salt should I buy?.
Salt mixes we stock: Tropic Marin Classic Sea Salt, Tropic Marin Pro-Reef Sea Salt and Brightwell NeoMarine Reef Salt Mix, or browse all salt & salinity gear.
Quick answer: what should a saltwater refractometer read?
Natural seawater averages 35 ppt salinity. On a seawater refractometer calibrated at 20°C, D-D says 35 ppt shows as 1.0266 specific gravity; Brightwell’s 35 ppt standard is stated as 1.0264 at 25°C. Most people round this to “about 1.026.”
| Setup | Salinity (ppt) | Source |
|---|---|---|
| Reef tank | 35–36 ppt | D-D refractometer instructions |
| Fish-only system | often kept down to 29 ppt | D-D refractometer instructions |
| Calibration standard | 35 ppt (SG 1.0264 at 25°C) | Brightwell Aquatics |
| RO, distilled or DI water | 0 ppt (zero point) | D-D and Milwaukee manuals |
Two ways to calibrate: zero it with RO, distilled or DI water (D-D’s and Milwaukee’s official method), or set it to 35 ppt with a 35 ppt calibration standard, which checks the scale right where a reef tank sits. Step-by-step instructions for both are below. Shop: D-D seawater refractometer · Milwaukee MA887 digital · Test kits.
What you’ll need
a handheld optical refractometer designed for natural seawater, not brine
$69.99
a digital alternative that Milwaukee says you calibrate with distilled or deionized water
From $199.00
Choose size
a 35 ppt reference solution, if you are calibrating to a known seawater standard rather than to zero with RO water
From $6.09
Choose sizeYou will also need RO, distilled, or deionized water for rinsing (and for a zero calibration on the D-D or Milwaukee), a soft cloth that will not scratch the prism, and a light source close to daylight if you are using the optical D-D unit. Rinse and dry the prism after every use so salt residue does not throw off the next reading.
Why a seawater scale is not the same as a brine scale
From D-D’s official instructions: almost all handheld salt refractometers in the hobby are designed and calibrated for brine (sodium chloride), not natural seawater. Other ions in seawater, including magnesium and calcium, change the refractive index. D-D’s example is that a 35 ppt solution of natural seawater has the same refractive index as a 36.5 ppt brine solution, so a brine refractometer would read 36.5 ppt when the seawater is actually 35 ppt. D-D’s seawater instrument is built for aquarium use, with a 0–40 ppt scale (2.5 times the resolution of a typical 0–100 ppt scale) and a copper body.
D-D Seawater RefractometerView product →
How ATC actually works
D-D’s instructions emphasize that salinity itself does not change with temperature, but a refractometer measures refractive index, which does. Automatic temperature compensation (ATC) uses a bimetal strip to move the scale as the instrument changes temperature. The small water sample equalizes to the instrument in seconds, so it is the refractometer’s temperature that matters, not the tank’s.
D-D units are calibrated at 20°C (68°F) and display specific gravity of seawater at that temperature. After a correct calibration at 20°C, D-D’s ATC is specified to work between 10°C and 30°C (50–86°F). Their FAQ also notes that if you take a reading immediately and again about 30 seconds later, the later reading is the one that has had time to equalize.
How D-D says to calibrate with RO water
From D-D’s official calibration steps:
- Let the instrument sit at 20°C (68°F) for 30 minutes so it is at its calibration temperature.
- Clean the glass prism and cover flap with pure or reverse-osmosis water so no salt residue remains.
- Rinse the supplied pipette several times in that same pure or RO water, then, holding the refractometer level, place a sample on the prism.
- Close the flap and wait 15 seconds for the sample to equalize (D-D’s FAQ describes waiting around 30 seconds). Point the prism toward a bright light close to daylight, and focus the eyepiece.
- Read where the view changes from blue to clear. If that boundary is not exactly on the zero mark, remove the protective cap and turn the adjustment screw with the supplied screwdriver until it is. The instrument is then calibrated.
D-D also says you can calibrate with a 35 ppt reference solution, but to be wary of solutions that do not state their calibration temperature, because they may be made for a different temperature than 20°C.
Using a 35 ppt calibration standard
Brightwell’s official standard is 35 ppt salinity, specific gravity 1.0264, refractive index 1.3394, and 53 µS/cm, all at 25°C.
Brightwell Calibration StandardView product →
Brightwell’s refractometer steps: shake the bottle well; clean the prism with fresh distilled water; dry it completely with a soft, clean, dry cloth; place several drops of the standard on the prism; wait 45 seconds; and calibrate to 35 ppt. Because Brightwell’s figures are stated at 25°C and D-D’s instrument is calibrated at 20°C, follow each maker’s own steps rather than mixing a 25°C standard with a 20°C zero-adjustment as if they were the same procedure.
For context, D-D says its refractometers display 1.0266 specific gravity at 35 ppt when calibrated at 20°C, which they describe as equivalent to 1.0264 on a hydrometer reading a 25°C sample with a 25°C / 77°F calibration.
How to read tank water on the D-D
- Rinse and dry the prism and cover with pure water after the last use, as D-D specifies.
- Rinse the pipette several times in the tank sample so old salt is not left in the dropper, then place a generous drop on the prism — D-D prefers enough water that the cover plate can push the excess away, rather than too little.
- Wait 15 seconds (or around 30 seconds, per D-D’s FAQ), then read salinity at the blue-to-clear boundary. One side of the scale is salinity; the other is specific gravity calculated at 20°C.
D-D recommends learning to think in salinity (ppt), because that is the scale calibrated against the refractive-index standard. They note that specific gravity depends on both the sample temperature and the instrument’s calibration temperature, and that a single 35 ppt sample can be described by a wide range of specific-gravity numbers for that reason. D-D also notes that ocean salinity varies, with a generally accepted global average of 35 ppt; that fish-only systems are often kept down to 29 ppt; that reef systems are normally kept at 35–36 ppt; and that the parameters printed for their H2Ocean Pro+ reef salt are quoted at 35.5 ppt. Keep your tank at the salinity that fits the animals you actually keep, and mix new saltwater to match — see Which salt should I buy?.
How Milwaukee says to use the MA887
The Milwaukee MA887 is a digital seawater refractometer. Milwaukee’s official product page lists accuracy of ±2 PSU, ±2 ppt, and ±0.002 S.G. (20/20), and ±0.3°C / ±0.5°F, with automatic temperature compensation. Milwaukee’s instruction manual says to calibrate with distilled or deionized water, daily before measurements, after a battery change, between a long series of readings, or after the environment has changed.
- Turn the meter on and wait until the display shows dashes.
- Fill the sample well completely with distilled or deionized water so the prism is covered. If the well is in intense light, Milwaukee says to shade it with your hand during calibration.
- Press ZERO. If no error appears, it is calibrated. Gently blot the water with a soft tissue — do not scratch the prism — and dry the well completely. Milwaukee says calibration is kept if you switch the meter off.
- To measure, start with a clean, completely dry prism, fill the well with sample, and wait about 1 minute if the sample temperature differs significantly from the instrument. Press READ. Milwaukee’s ATC is specified from 0–40°C (32–104°F); the ATC tag blinks and compensation is disabled outside that range.
- Blot the sample, rinse with distilled or deionized water, and dry before the next reading.
Use the salinity or specific-gravity unit you have chosen in the meter’s setup, and record the unit with the number. For a broader testing plan, see Testing your water.